IEC60068 Universal Spring Impact Hammer Measurement Calibration Test/Testing Machine
1. Product Description
It is designed and manufactured according to the standard requirements of IEC60068, suitable to the calibration for spring hammer with impact energy less than 2J. calibration device main structures includes: bearing, pointer, pendulum, releasing mechanism, releasing substrate and divided circle.
2. Standards
IEC60068
3. Technical Parameters
A:
1. Compare the pendulum quality to the dropping calculated energy value, so calculate the impact hammer actual value. |
2. Spring hammer with adjusted height put in the releasing substrate, the releasing mechanism of calibration device operated for three time, the striking element is turn to different scale interval for each time, the standard value of three time readings is regard as the actual value of spring impact hammer. |
3. Using unmarked meter regulator to examine the impactor. |
4. Can do the examine test for spring impactor with 0.1-2J. |
B:
Impact element |
250g |
Hammer diameter |
10mm |
A set of portal frame fixed on the working bench, install four hanging ropes |
vertical height is 2000mm |
Impact element releasing height |
408mm |
With height scale |
|