Model | According to the standard | Average volume | Appearance size (mm) | Weight |
Technical requirements | Configuration |
SH5101 Glow-wire tester |
IEC60335 and IEC60695 | 0.1 | 710(Width) x365 (Depths)x760(Height) | 30 | 1.The temperature of incandescent wire could range from 500ºC to 1000ºC, with a temperature tolerance for error of ±10ºC between 500ºC and 750ºC, and ±15ºC between 750 and 1000ºC. The accuracy of temperature measurement instrument is± 0.5 class. The thermocouple for measuring the temperature is armored with the nickel-chromium (K) wire of 0.5-0.55mm. 2.The incandescent wire imposed on the samples the power of 0.8N to 1.2N, and the restrict indentation depth is at least 7 mm. 3.The time of activation (Ta) on the specimen by the incandescent wire ranges from 0 to 99 minutes and 99 seconds. Normally, according to the standard requirements, set the time to 30 seconds, and a timer will be employed to count the time of ignition(Ti) and the time of extinguishments, with the time error of ±1S. 4,The operation of the equipment is simple, the temperature measurement system is reliable and has automatic temperature compensation function. |
0.5 Armored Thermocouple |